Detect trace level TBC in 1,3 Butadiene
Current method ASTM D1157 determines TBC traces in 1,3 Butadiene with a detection limit of 50 mg/kg and requires evaporation of large volumes of Butadiene. Our new Application Note presents an alternative approach developed by Da Vinci Laboratory Solutions, using a GC system equipped with a Liquefied Gas Injector: the Butadiene Analyzer.
Safe, Accurate and Reliable GC Solution
Tests of TBC in 1,3 Butadiene using the Butadiene Analyzer show a consistent linear response, no carry over, detection limits below 1 mg/kg, and RSDs under 5%. This approach also eliminates the need to evaporate large sample volumes, improving analyst safety while ensuring accurate and reliable quantification.
The analysis results indicate that the DVLS Butadiene Analyzer is a safer and more reliable alternative to the ASTM D1157 method, offering improved performance in terms of sensitivity and reproducibility.
Download the Application Note to learn more about the determination of TBC Linearity and Carry-over in 1,3-Butadiene with the DVLS Butadiene Analyzer.